a.
Procedure (to Attach)
Figure 2-2 Attach / Remove Pressure connections
Table 2-1
Step
Procedure
1
Remove the adaptor
2
Use an applicable seal for the pressure connection:
a. NPT type: Use an applicable sealant on the thread.
b. BSP (parallel) type: We recommend a bonded seal at
the bottom.
c. BSP (parallel type, 100 bar (1500 psi) or less: a bonded
seal at the top is permitted.
3
Attach the adaptor to the device; if necessary use one of
the alternative adaptors listed in
Section 1.10
(Accessories), then tighten to the applicable torque.
b. (G)/BSPP
P≥100 bar
c. (G)/BSPP
P≤100 bar
Pressure Port
a. NPT
1
2
3