ANALYSIS FUNCTION
4.3
Set Input channel parameters (See
Section 4.2
), and return to
CHANNEL SETUP screen.
Select the Start button
.
The Analysis window displays the following:
•
The deviation of each Input channel from the ideal transfer
characteristic.
•
A tolerance limit test icon
Pass
(within tolerance test limit)
Fail
(outside tolerance test limit)
To check the full range of the device do the following:
1.
Step the Reference signal value through its range.
2.
Check the Analysis window at each step.
3.
If the Reference is sourced from the calibrator, move to the
channel window to change the Reference value.
4.
Return to the Analysis window.
5.
When the analysis is complete exit the window by selecting exit
.
RUN PROCEDURE
4.4
The purpose of the Run procedure is to do calibration procedures
which have been downloaded from 4 Sight software. The 4 Sight
calibration procedures contain all the values to calibrate a device
under test (test points, ramp time).
The same calibration procedure can be used for all the applicable
devices under test. To use the Run Procedure function, the following
items are required:
•
A copy of the 4 Sight Calibration Software.
•
Standard USB lead (as supplied).