User Manual
Druck DPI612
Page 90 of 105
[EN] English – 109M4017 Revision -
Table 5-3 Current (source) error limits
Applied
mA
Calibrator
uncertainty
(mA)
Permitted
DPI612 error
(mA)
0.2
0.00008
0.0013
6
0.00023
0.0017
12
0.00044
0.0020
18
0.0065
0.0024
24
0.0012
0.0028
PROCEDURES: DC mV/Volts (measure)
5.7
Do the procedure as follows:
1.
Connect the applicable calibration equipment (See
2.
Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3.
Use the calibration menu (
See Section 5.4
) to do a three-point
calibration.
(-FS, Zero and +FS) for the applicable set of ranges:
mV (measure) ranges
Volts (measure) ranges
200 mV
20 V
2000 mV
30 V
4.
Check the calibration is correct:
•
Select the applicable Millivolts or Voltage (measure) function.
•
Apply the input values that are applicable to the Calibration:
mV: -2000, -1000, -200, -100, 0 (short circuit)
mV: 0, 100, 200, 1000, 2000
Volts (V): -30, -21, -20, -10, -5, 0 (short circuit)
Volts (V): 0, 5, 10, 20, 21, 30.
•
Check that the error is within limits (
See Table 5-4