User Manual
Druck DPI612
Page 91 of 105
[EN] English – 109M4017 Revision -
Table 5-4 Millivolts (measure) error limits
Applied
mV
Calibrator
uncertainty
(mV)
Permitted
DPI612 error
(mV)
±2000
0.051
0.1280
±1000
0.040
0.0940
±200
0.051
0.0148
±100
0.0040
0.0110
0 (short circuit)
0.0036
0.0070
Table 5-5 Voltage (measure) error limits
Applied
V
Calibrator
uncertainty
(V)
Permitted
DPI612 error
(V)
±30
0.00052
0.00180
±21
0.00040
0.00154
±20
0.00031
0.00118
±10
0.00016
0.00089
±5
0.00008
0.00075
0
0.000024
0.00060
PROCEDURES: DC Volts (source)
5.8
Do the procedure as follows:
1.
Connect the applicable calibration equipment (See
2.
Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3.
Use the calibration menu (See
Section 5.4
) to do a two-point
calibration for the applicable range:
Volts (source) range
10 V
24 V