User Manual
Druck DPI612
Page 88 of 105
[EN] English – 109M4017 Revision -
Before starting the calibration, check the time and date on the
instrument are correct.
To do a calibration on a measure or source function, use the
advanced menu option.
Select:
Enter the calibration PIN: 4321
Select the
button.
Select:
Then select a function and start the calibration:
1.
Select channel
2.
Select function
3.
Select range (if applicable)
4.
Follow on-screen instructions
When the calibration is complete, set the next calibration date.
PROCEDURES: CURRENT (measure)
5.5
Do the procedure as follows:
1.
Connect the applicable calibration equipment (
See Table 5-1
2.
Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3.
Use the calibration menu (
See Section 5.4
) to do a three-point
calibration (-FS, Zero and +FS) for each range: 20 mA and 55 mA.
4.
Check the calibration is correct.
•
Select the applicable Current (measure) function.
DASHBOARD >>
SETTINGS
>> ADVANCED
PERFORM CALIBRATION