[EN] English – 109M4017 Revision -
4.
Check the calibration is correct:
•
Select the applicable Voltage (10V) or Current (24V)
(measure) function (
See Section 5.7
10V
24V
•
Check the error is within limits (
See Table 5-6
Table 5-6 Voltage (source) error limits
Source
V
Calibrator
uncertainty
(V)
Permitted
DPI612 error
(V)
10
0.0001
0.0015
24
0.0004
0.0036
PROCEDURES: PRESSURE INDICATOR
5.9
Do the procedure as follows:
1.
Connect the instrument to the pressure standard.
2.
Let the equipment get to a stable temperature (minimum: 60
minutes since the last power on).
3.
Use the calibration menu (
See Section 5.4
) to do a two-point
calibration.
•
Zero and +FS for gauge sensors.
4.
Check the calibration is correct:
•
Select the applicable pressure function.
•
Apply the following pressure values.
0, 20, 40, 60, 80, 100, 80, 60, 40, 20 (%FS).
•
Apply the following pressure values.
-200, -400, -600, -800, -600, -400, -200, 0 mbar.
•
Check the error is within limits
From the datasheet use the values in the Accuracy column.