Druck ADTS 2XX User Manual
20
K0451 Issue No. 1
Altitude Correction
An altitude correction must be made to allow for the
difference in height between the reference level of the
ADTS 2XX and the reference level of the aircraft’s
altitude sensors. Select the Instrument Altitude
Correction and, using the numeric keys, enter the
correction value (see Fig 2-2).
RS232 Settings
The RS232 interface enables the unit to communicate
with a PC. The parameters for RS232
communications are applied through the set-up
menu.
Baud rate
-
600, 1200, 2400, 9600
Data bits
-
8
Stop bits
-
1,2
Parity
-
none, even, odd
Auto leak
Auto leak automatically regains control if the leak rate becomes too high for any of the control channels. Auto
leak operates at preset rates of 3000ft/min and 300knots/min. This facility can be switched off and on in set-
up menu.
Pressure Monitor
Set-up selection can be made between the display
showing the “REMOTE HAND TERMINAL OPERATING”
and the pressure monitor screen. The pressure monitor
shows the current pressures of all three channels in the
selected units of pressure measurement.