37
Druck ADTS 2XX User Manual
K0451 Issue No. 1
UNITS:PRES
Command Syntax
UNITs:PRESsure <units>
Parameter Type:
DISCRETE
Parameter Range:
MBAR, INHG, PSI, MMHG, KPA, HPA
Function:
To set the pressure units used for SCPI commands and queries. The display
only shows these units when SOUR:PRES or SOUR:RATE is sent.
Conditions:
None
Query Syntax
UNITs:PRESsure?
Parameter Type:
None
Parameter Range:
Returned Data
<units>
Returned Data Type
DISCRETE
Returned Data Range:
MBAR, INHG, PSI, MMHG, KPA, HPA
Function:
To query the SCPI pressure units.
Conditions:
None