![background image](http://html.mh-extra.com/html/ge/druck-adts-2-series/druck-adts-2-series_user-manual_132649042.webp)
Druck ADTS 2XX User Manual
30
K0451 Issue No. 1
STAT
This group of commands monitor and control the status bytes that show the status of the ADTS 2XX.
STAT:OPER:CON
Command Syntax
N/A
Parameter Type:
Parameter Range:
Function:
Conditions:
Query Syntax
STATus:OPERation:CONdition?
Parameter Type:
None
Parameter Range:
Returned Data:
<status>
Returned Data Type:
INTEGER
Returned Data Range:
0-32767
Function:
To query the status of the ADTS. The table shows the function of each bit.
Conditions:
None
Bit
Data
0
Reserved – returns 0
1
Ps1 controlling and at set-point
2
At ground
3
Ps2 controlling and at set-point
4
Pt controlling and at set-point
5
Reserved – returns 0
6
Reserved – returns 0
7
Reserved – returns 0
8
Reserved – returns 0
9
Reserved – returns 0
10
Reserved – returns 0
11
Reserved – returns 0
12
Reserved – returns 0
13
Reserved – returns 0
14
Reserved – returns 0
15
Reserved – returns 0
Summary of Contents for Druck ADTS 2 Series
Page 1: ...GE Sensing Druck Air Data Test Systems Three channel Test Set User Manual K0451 XX ...
Page 2: ... General Electric Company All rights reserved ...
Page 56: ...Druck ADTS 2XX User Manual 44 K0451 Issue No 1 INTENTIONALLY LEFT BLANK ...
Page 66: ...Druck ADTS 2XX User Manual 54 K0451 Issue No 1 INTENTIONALLY LEFT BLANK ...