6.2 Air Data Specifications
Note:
All range values at maximum limits.
Ps 1 Port
Pressure Mode
Control Range: ................................................................................................................... 35
(1)
to 1355 mbar a
Measure Range: .................................................................................................................. 27.6 to 1355 mbar a
Resolution: ....................................................................................................................................................0.01mbar
Accuracy: .....................................................................................................................................................±0.1 mbar
Rate Range:............................................................................................................................0 to 1000 mbar/min
Rate Accuracy: .................................................................................................................. ±1% of aim (average)
Aeronautical Mode
Control Range: ........................................................................................................................ -2000 to 60,000 ft.
Measure Range: ..................................................................................................................... -3000 to 100,000ft.
Resolution: ................................................................................................................................................................ 1 ft.
Accuracy: ......................................................................................................................................±3 ft. (@sea level)
..............................................................................................................................................±7 ft. (@30,000 ft)
...........................................................................................................................................±23 ft. (@55,000 ft)
Rate Range:.................................................................................................................................0 to 40,000 ft/min
Rate Accuracy: .................................................................................................................. ±1% of aim (average)
Ps 2 Port
Pressure Mode
Control Range: ................................................................................................................... 35
(1)
to 1355 mbar a
Measure Range: .................................................................................................................. 27.6 to 1355 mbar a
Resolution: ....................................................................................................................................................0.01mbar
Accuracy: .....................................................................................................................................................±0.1 mbar
Rate Range:............................................................................................................................0 to 1000 mbar/min
Rate Accuracy: .................................................................................................................. ±1% of aim (average)
Aeronautical Mode
Control Range: ........................................................................................................................ -2000 to 60,000 ft.
Measure Range: ..................................................................................................................... -3000 to 100,000ft.
Resolution: ................................................................................................................................................................ 1 ft.
Accuracy: ......................................................................................................................................±3 ft. (@sea level)
..............................................................................................................................................±7 ft. (@30,000 ft)
...........................................................................................................................................±23 ft. (@55,000 ft)
Rate Range:.................................................................................................................................0 to 40,000 ft/min
Rate Accuracy: .................................................................................................................. ±1% of aim (average)
(1)
limited only by vacuum pump/system leaks
Summary of Contents for Druck ADTS 2 Series
Page 1: ...GE Sensing Druck Air Data Test Systems Three channel Test Set User Manual K0451 XX ...
Page 2: ... General Electric Company All rights reserved ...
Page 56: ...Druck ADTS 2XX User Manual 44 K0451 Issue No 1 INTENTIONALLY LEFT BLANK ...
Page 66: ...Druck ADTS 2XX User Manual 54 K0451 Issue No 1 INTENTIONALLY LEFT BLANK ...