245
"Controller has an invalid EEPROM
archive"
Invalid/corrupted data - recovery
in progress
Check/correct calibration
247
"Handset has an invalid EEPROM
archive"
Invalid/corrupted data - recovery
in progress
250
"Vacuum pump processor has
failed"
Internal fault
Return unit to repair depot
251
"Vacuum pump lead is
disconnected"
-
Return unit to repair depot
252
"Vacuum pump transistor is faulty
or drive inhibited"
Internal fault
Return unit to repair depot
No.
Message
Probable Cause
Action
253
"Vacuum pump has stalled"
Internal mechanical fault
Restart system, if fault repeats -
return unit to repair depot
254
"Pressure pump processor has
failed"
Internal fault
Return unit to repair depot
255
"Pressure pump lead is
disconnected"
-
Return unit to repair depot
256
Pressure pump transistor is faulty
or drive inhibited"
Internal fault
Return unit to repair depot
257
Pressure pump has stalled"
Internal mechanical fault
Restart system, if fault repeats -
return unit to repair depot
260
Connected handset has failed to
respond. Disconnecting power."
Handset bad connection or
comms/power supply fault
Check connections and restart, if
fault repeats - return unit to
repair depot
Summary of Contents for Druck ADTS 2 Series
Page 1: ...GE Sensing Druck Air Data Test Systems Three channel Test Set User Manual K0451 XX ...
Page 2: ... General Electric Company All rights reserved ...
Page 56: ...Druck ADTS 2XX User Manual 44 K0451 Issue No 1 INTENTIONALLY LEFT BLANK ...
Page 66: ...Druck ADTS 2XX User Manual 54 K0451 Issue No 1 INTENTIONALLY LEFT BLANK ...