![background image](http://html.mh-extra.com/html/ge/druck-adts-2-series/druck-adts-2-series_user-manual_132649044.webp)
Druck ADTS 2XX User Manual
32
K0451 Issue No. 1
SOUR:PRES
Command Syntax
SOURce:PRESsure <parameter>, <aim>, <aim>
Parameter Type:
DISCRETE, REAL
Parameter Range:
ALT | CAS | MACH | PS | PT | QC
Static Channels:
SOUR:PRES <parameter>,<Ps1 Aim>,<Ps2 Aim>
Pitot Channel:
SOUR:PRES <parameter>,<Pt Aim>
Function:
To command the selected parameter to a new aim value.
Conditions:
Controllers must be ON before using this command.
Query Syntax
SOURce:PRESsure? <parameter>
Parameter Type:
DISCRETE
Parameter Range:
ALT | CAS | MACH | PS | PT | QC
Units:
As set UNIT command
Returned Static Channels:
<Ps1 Aim>,<Ps2 Aim>
Returned Pitot Channel:
<Pt Aim>, *
Function:
To query the last commanded aim values
Conditions:
See current limit settings
Summary of Contents for Druck ADTS 2 Series
Page 1: ...GE Sensing Druck Air Data Test Systems Three channel Test Set User Manual K0451 XX ...
Page 2: ... General Electric Company All rights reserved ...
Page 56: ...Druck ADTS 2XX User Manual 44 K0451 Issue No 1 INTENTIONALLY LEFT BLANK ...