![background image](http://html.mh-extra.com/html/ge/druck-adts-2-series/druck-adts-2-series_user-manual_132649070.webp)
Druck ADTS 2XX User Manual
58
K0451 Issue No. 1
Name: “STANDARD”
MIN ALT
-2000 ft
MAX ALT
60,000 ft
MAX ROC
10,000 ft/min
MIN CAS
0.0 knots
MAX CAS
650 knots
MAX MACH
1.732 Mach
MIN Ps1/Ps2
65.00 mbar (1.9194 inHg)
MAX Ps1/Ps2
1088.60 mbar (32.1463 inHg)
MIN Qc
0.00 mbar (0.0000 inHg)
MAX Qc
866.00 mbar (25.5730 inHg)
MAX Rate Ps1/Ps1
200.00 mbar/min (5.9060 inHg/min)
MAX Rate Qc
200.00 mbar/min (5.9060 inHg/min)
ALT CORRECTION
100 ft
ARINC LIMITS
OFF
MAX Ps1 to Ps2 Diff.
84.66 mbar (2.5 inHg)
Name: “CIVIL”
MIN ALT
-1000 ft
MAX ALT
50,000 ft
MAX ROC
6,000 ft/min
MIN CAS
0.0 knots
MAX CAS
450 knots
MAX MACH
1.0 Mach
MIN Ps1/Ps2
115.97 mbar (3.4246 inHg)
MAX Ps1/Ps2
1050.41 mbar (31.0185 inHg)
MIN Qc
0.00 mbar (0.0000 inHg)
MAX Qc
368.01 mbar (10.8673 inHg)
MAX Rate Ps1/Ps1
109.85 mbar/min (3.2439 inHg/min)
MAX Rate Qc
109.85 mbar/min (3.2439 inHg/min)
ALT CORRECTION
100 ft
ARINC LIMITS
OFF
MAX Ps1 to Ps2 Diff.
84.66 mbar (2.5 inHg)