TROUBLESHOOTING and
FREQUENTLY ASKED QUESTIONS
ECG
Problem:
Why is the monitor alarming for asystole, bradycardia,
pause or inaccurate heart rate when a visible QRS is
present?
Answer:
The monitor may not be seeing sufficient QRS
amplitude in all analyzed leads. Multiple leads
(I, II, III and V) are used for arrhythmia processing.
For best results, a 1.0 mV amplitude in all analyzed leads
is recommended.
Intervention:
■
Check the ECG signal acquired from the patient:
•
Review the
VIEW ALL ECG
option to assess the
amplitude of the QRS complexes. A minimum of
0.5 mV amplitude in all analyzed ECG waveforms
at normal size (1x) is required for QRS detection.
For best results, a 1.0 mV amplitude in all
analyzed leads is recommended. Amplitude is
viewed in one direction (positive or negative).
For borderline signals, validate the ECG
waveform on a graph.
•
If the amplitude is low in any analyzed leads,
perform skin preparation and adjust electrode
placement. When adjusting placement, utilize
fresh electrodes.
–
Increasing the amplitude >2X when using the
ECG SIZE
option is for the clinicians viewing
only. It will not affect the ECG analysis.
–
It may be beneficial to move the V-Lead
electrode (chest lead) to an alternate precordial
electrode placement to improve detection.
(continued on back of card)
Summary of Contents for DASH 3000 V5
Page 30: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 42: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 76: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 90: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM Optional Equipment ...
Page 104: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...