The Manual Mode overrides
the Auto Mode and requires
additional steps at the
monitor to complete the
measurement. Under certain
patient conditions (valvular
disease or respiratory
variation in PA reading), the
monitor is unable to detect a
change in the waveform.
NOTE:
The monitor must detect a 30% change in the
waveform to measure a wedge. If the waveform does not
change accordingly, the Wedge Processing message will not
appear and you must use the manual mode for wedge
measurements.
•
To Complete a Manual PA
Wedge:
–
Select
PA
.
–
Select
PA INSERT/
WEDGE
.
–
Verify that the
MODE:MANUAL
is
displayed.
–
Inflate the balloon.
–
Watch PA waveform for
wedging.
–
Deflate the balloon.
–
Select
REVIEW WEDGE
.
–
The PA Wedge Review
Information window is
displayed.
–
Select
MOVE WEDGE
CURSOR
.
–
Turn the Trim Knob to
move the cursor to the
desired position and press
to select.
PA Insert/Wedge
PA Wedge Review
–
Press the
GRAPH GO/
STOP
button on the
monitor to print window.
–
Select
Main Menu
to exit.
NOTE:
The last PAW with a time stamp will be displayed in the
PA window.
Summary of Contents for DASH 3000 V5
Page 30: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 42: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 76: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 90: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM Optional Equipment ...
Page 104: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...