Relearn:
During ECG
monitoring, it may be necessary
to relearn the ECG waveform if
a change in the patient’s normal
ECG pattern has occurred, or
the electrode placement has
changed.
•
A change in the ECG pattern
could result in:
–
Incorrect arrhythmia calls.
–
Loss of ST measurement.
–
Inaccurate heart rate
detection.
•
To Relearn the ECG
Waveform:
–
Select the ECG parameter
window.
–
Select
RELEARN
.
–
The ECG parameter
window will replace the
HR
with an “
X
” and the
message “
Learning
” will
appear above the display
lead.
–
Select
MAIN MENU
to
exit.
■
Turning the ECG Parameter
Off or On:
See monitor setup
for details.