PA Insert/Wedge is a feature
found in the PA Invasive
Pressures Menu. There are two
modes for doing a PA Wedge
measurement: Auto and
Manual.
■
The
Auto Mode
allows a PA
Wedge measurement to be
performed without having to
touch the monitor after
starting the program. The
Manual Mode defaults when
the PA Insert/Wedge menu
option is selected.
NOTE:
The monitor must detect a 30% change in the
waveform to measure a wedge. If the waveform does not
change accordingly, the Wedge Processing message will not
appear and you must use the manual mode for wedge
measurements.
•
To Complete an Auto PA
Wedge:
–
Select
PA
.
–
Select
PA INSERT/
WEDGE
.
–
Select
MODE:MANUAL
(this changes the mode to
Auto).
–
An I
NFLATE BALLOON
message appears in the PA
parameter window.
–
Deflate the balloon.
–
PA Wedge Review
Information window is
automatically displayed.
–
To change the PA value,
select
MOVE WEDGE
CURSOR
.
–
Turn the Trim Knob to
move the cursor to the
desired position and press
to select.
–
Press the
GRAPH GO/
STOP
button on the
monitor to print window.
–
Select
MAIN MENU
to
exit.
NOTE:
The last PAW with a time stamp will be displayed in the
PA parameter window.
PA Insert/Wedge
PA Wedge Review
Summary of Contents for DASH 3000 V5
Page 30: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 42: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 76: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 90: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM Optional Equipment ...
Page 104: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...