Alarms
Problem:
Why would the monitor
alarm
for a low heart rate or
high heart rate, but not store in Alarm History?
Answer:
Parameter alarms (i.e., limit violations) do not store in
Alarm History. The Tachy and Brady arrhythmia alarm
categories must be set in advisory or above to store in the
Alarm History.
Intervention:
■
Review Arrhythmia Alarm Levels in the Alarm
Control menu.
■
Adjust the Tachy and Brady alarms to Crisis,
Warning or Advisory level.
Problem:
Why is the monitor not calling
PAUSE
when the patient
is experiencing frequent pauses?
Answer:
The current criteria for the
PAUSE
may not be
exceeded. The
PAUSE
alarm in the Dash monitor allows
the clinician to define the pause alarm intervals for
individual patients.
Intervention:
■
Review the
Arrhythmia Alarm Levels
in the Alarm
Control menu to ensure it is listed in Crisis, Warning
or Advisory.
■
Review the
BEAT PAUSE INTERVAL
in the ECG
menu to determine the current crisis.
■
Adjust the
BEAT PAUSE INTERVAL
. Selections
available are from
one second
to
Five seconds
in
1
/
2
second intervals.
TROUBLESHOOTING and
FREQUENTLY ASKED QUESTIONS
Summary of Contents for DASH 3000 V5
Page 30: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 42: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 76: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 90: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM Optional Equipment ...
Page 104: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...