Detect Pace:
Turns pacemaker
detection On/Off.
•
Two different options of
pacemaker modes are
available. The clinician must
judge which mode is best for
each patient.
Pace 2
is the
recommended pacemaker
detection mode.
Pace 1
should be used if Pace 2 does
not adequately detect
pacemaker spikes.
•
To Activate or Deactivate the
Pacemaker Mode:
–
Select
ECG
.
–
Select
DETECT PACE
.
–
Turn the Trim Knob in
front of the desired
pacemaker mode and press
to select.
–
Select
MAIN MENU
to
exit.
NOTE:
A “
P
” appears in the ECG parameter window when pace
detection is enabled. An “
*
” will appear if a paced beat is
detected.
■
Pace 2 Mode:
Analyzes
waveforms with the added
capability of minimizing the
chance of counting severe
residual pacemaker energy as a
QRS complex. Pace 2 may not
adequately detect all QRS
morphologies. Arrhythmia calls
such as
Asystole
or
Pause
may
be made with heart rate
identified as less than actual.
NOTE:
Pacemaker patients should be kept under close
observation.
■
Pace 1 Mode:
Analyzes the
presence of pacer spikes,
assesses the waveform for
residual pacemaker energy and
determines the presence of an
R-wave following the pacemaker
spike. If an event occurs during
the first few milliseconds
following the spike, it will be
counted as a paced spike.
Summary of Contents for DASH 3000 V5
Page 30: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 42: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 76: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...
Page 90: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM Optional Equipment ...
Page 104: ...TAB TRIM COVER NON TAB TRIM TAB TRIM COVER NON TAB TRIM ...