250cx Series Maternal/Fetal Monitor
9-11
2036946-001
Maternal Pulse Oximetry Monitoring: Module and Probe Compatibility
MSpO
2
Pulse Beat Audio
If MSpO
2
is selected as the MHR/P source, each pulse beat is indicated with a
“beep”: the pitch of the beep will vary according to the saturation value; the pitch
rises as the saturation value increases; the pitch lowers as the saturation value
decreases. If MECG is selected as the MHR/P source, the MECG audio
plink
is used
instead; it will not vary in pitch. NIBP cannot be chosen as the MHR/P source.
(Refer to Chapter 7, “Maternal Heart/Pulse Rate Monitoring” for more information.)
The MSpO
2
Waveform
When MSpO
2
monitoring is employed, the MSpO
2
pulsatile (plethysmograph)
waveform can be displayed and printed. Refer to Chapter 14, “Waveforms” for more
information.
When enabled, the MSpO
2
trend prints in the bottom grid as a grey trace annotated
by MSpO
2
. Values are printed on the annotation area preceded by an outlined
diamond
which marks the time of the reading.
Module and Probe Compatibility
Ohmeda, Masimo, and Nellcor pulse oximetry parameters are calibrated to display
functional saturation. Other manufacturer’s pulse oximetry monitors may be
calibrated to display fractional saturation.
250cx Series Monitors with Masimo Technology are compatible only with
Masimo LNOP and LNCS sensors. For additional information, refer to
Directions for Use supplied with sensor.
250cx Series Monitors with Ohmeda Technology are compatible only with
Ohmeda sensors. For additional information, refer to Directions for
Use supplied with sensor.
250cx Series Monitors with Nellcor Technology are compatible only
with Nellcor OxiMax sensors. For additional information, refer to
Directions for Use supplied with sensor.
IMPORTANT:
Use only Masimo LNOP oximetry sensors with the Masimo
Technology, Ohmeda sensors with the Ohmeda Technology, and Nellcor
sensors with the Nellcor Technology. Other sensors may result in unpredictable
performance.
The MSpO
2
cable should plug into the monitor’s
MSpO
2
connector easily and
securely. Do not use excessive force to connect the cable. If the MSpO
2
cable does
not easily fit into the
MSpO
2
connector on the monitor, it is likely that you are using
an incorrect cable.
IMPORTANT:
It is possible to connect the wrong MSpO
2
cable and/or
sensor to the monitor. If this happens, the MSpO
2
parameter will not work
(Ohmeda and Masimo) or will cause an error (Nellcor error:
SENSOR
). Be sure
to check the type of MSpO
2
technology your monitor contains (the label next to
the lower, right-hand side of the display) and use the corresponding cables and
sensors for that technology.
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......