11-10
250cx Series Maternal/Fetal Monitor
Revision C
2036946-001
Recorder Modes: Summary of Annotations
Multiple Annotations Example
Summary of Annotations
Summary of Annotations
Annotation
Explanation
Time and Date
(Example: 10:40 12 AUG 97)
Time and date are both printed on the bottom annotation line 20 seconds after
the recorder is turned on and when the date changes after midnight.
A time stamp automatically prints approximately every 10 minutes—at the 10-
minute mark. For example: 10:50, 11:00, 11:10, 11:20, 11:30, etc. If the bottom
annotation line is being used to print another annotation, the time stamp is
delayed. For example: 10:50, 11:00, 11:12, 11:20, 11:30, etc. In this example, the
11:10 date stamp was delayed until 11:12.
The time and/or date also prints whenever it is changed via the General Setup
screen.
SET TIME/DATE
If the monitor senses a clock circuit fault, when the recorder is turned on, this
message replaces the normal time/date stamp. The message reprints every 10
minutes, at the 10-minute mark, until the clock is reset.
TEST: ARE ALL DOTS PRINTED?
This annotation prints across the width of the top strip chart grid when you press
the Test button. The message reminds you to check for a continuous unbroken
line of recorder dots.
This icons prints prior to the FHR trend source annotations if the FHR alarms are
enabled. The FHR alarm option is enabled/disabled via the password-protected
Install Options service screen.
US or FECG
The trend source prints on the bottom annotation line by the following rules:
All trend sources print 20 seconds after the recorder is turned on, including
inoperative modes.
All trend sources print every 30 minutes.
If a mode change occurs, only those trend sources belonging to the
corresponding group print. If any top grid trend source changes, all top grid
active trend sources are printed. If the UA mode changes, the active UA trend
source is printed. A mode change is defined as: switching connectors;
connecting to a front panel receptacle; disconnecting from a front panel
receptacle; or enabling/disabling a trend on a setup menu.
US or US2
MECG or MSpO
2
P
UA
External FSpO
2
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......