7-4
250cx Series Maternal/Fetal Monitor
Revision C
2036946-001
Maternal Heart/Pulse Rate Monitoring: MHR/P Setup Screen
MHR/P Setup Screen
Select the mode title softkey—
MECG
or
Pulse
—to access the MHR/P Setup screen.
(Refer to the following figure.)
NOTES
The figure below provides an example of MECG selected as the MHR/P
source, as indicated by the
MECG
mode title. When either MSpO
2
or NIBP
are selected as the MHR/P source, the mode title changes to
Pulse
.
The
Lead
source and
Pacer
fields apply to MECG only.
MHR/P Setup Screen
Source
This field selects the MHR/P source. When
Auto
is selected, the monitor checks for
parameter availability and use in the following order: MECG, MSpO
2
, then NIBP. If
a source is not available, the next available source is automatically selected.
IMPORTANT
WAVEFORM—The MHR/P
Source
field is independent of the
waveform selected on the normal operating screen. For example,
you can select MECG as the MHR/P source yet display the
MSpO
2
plethysmograph waveform. Or, you can select MSpO
2
as
the source and display MECG as the waveform.
Alarms
High
Low
MHR/P:
120
50
bpm
Alarm Volume 5
165
172
30
US
FECG
TOCO
MHR/P Setup
89
MECG
Source:
HR/PR Trace:
Volume: 5
MECG
Lead: II
Pacer: Off
35-120
On
Off
I, II, III
On
Off
Auto
MECG
MSpO
2
0
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
Auto
Off
Exit
100-250
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......