16-4
250cx Series Maternal/Fetal Monitor
Revision C
2036946-001
Troubleshooting: Ultrasound Troubleshooting
Ultrasound Troubleshooting
Ultrasound Troubleshooting
Problem
Probable Cause
Possible Solution
Ultrasound not functioning properly.
Transducer not properly connected to
monitor.
Transducer placement.
Too little gel applied to transducer.
Defective transducer.
Active fetus or mother. Fetal arrhythmia or
hiccups. Extreme maternal obesity.
No signal.
Service required.
Ensure that transducer is securely
attached to monitor.
Wait before moving transducer; FHR
often returns. Reposition transducer.
Apply more gel.
Replace transducer.
Use alternate technique.
Auscultate FHR.
Call Biomedical Engineering
Department.
Check maternal pulse to ensure
monitoring of FHR.
Static noise on ultrasound.
Active fetus.
Environmental noise.
Maternal movement.
Defective transducer.
Reposition transducer.
Keep sheets and gown off transducer.
Do not hold transducer with hand.
Use alternate monitoring mode.
Replace transducer.
Rate on FHR area of display and FHR
trend on strip chart paper do not correlate.
Monitor is set for 30 bpm/cm vertical scale
and 20 bpm/cm vertical scale strip chart
recorder paper is being used (or vice versa).
Call Biomedical Engineering Department.
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......