10-8
250cx Series Maternal/Fetal Monitor
Revision C
2036946-001
Alarms: Maternal Alarms
Active Signal Quality Alarm
The following signal quality alarms are indicated both visually and audibly. The
audio alarm consists of alternating high/low tones. The visual indication varies
according to the alarm:
Asystole:
Dashes “– – –” display in place of the MHR/P numeric.
MECG Leads Off:
Dashes “– – –” display in place of the MHR/P numeric.
The following message displays in the MHR/P area:
MECG LEADS OFF
. During
this type of alarm, the MHR/P source automatically switches to the next
available parameter (MSpO
2
then NIBP). As soon as the alarm condition is
resolved and the MECG signal is detected, and the monitor resumes using
MECG as the MHR/P source.
NIBP System Problem:
When there is a malfunction with the NIBP parameter,
cuff, or air hoses, the monitor will be unable to make a determination. During
this type of alarm, one of the following messages displays in the
NIBP
area:
CHECK CUFF
,
OVERPRESSURE
,
COMM
,
MOTION
,
WEAK SIGNAL
, or
REPAIR
.
Refer to Chapter 16, “Troubleshooting”, for more information on these
messages.
MSpO
2
System Problem:
When there is a malfunction with the monitor’s
built-in MSpO
2
parameter, one of the following messages displays in the
MSpO
2
area:
COMM
,
REPAIR
or
SENSOR
(Nellcor only). Refer to Chapter 16,
“Troubleshooting”, for more information.
MSpO
2
Disconnect:
An MSpO
2
disconnect alarm occurs if: the MSpO
2
intermediate cable is disconnected from the monitor, the sensor assembly is
disconnected from the intermediate cable, or the sensor or cable have a broken
wire. Dashes “– – –” display in place of the MSpO
2
numeric.
Resolved Signal Quality Alarm
Resolved signal quality alarms behave like most other 250cx Series alarms. As soon
as an alarm condition is resolved, the visual and audible indications automatically
disappear.
Silencing a Maternal Audio Alarm
Press the
Alarm Silence
button to cancel the audio; however, the visual indication
remains until the condition is resolved.
For MECG and MSpO
2
, you can only
temporarily
silence the audio portion of the
alarm. If the alarm condition remains, after expiration of the re-alarm time
configured on the Master Alarm Setup screen, the audio alarm is re-issued.
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......