10-6
250cx Series Maternal/Fetal Monitor
Revision C
2036946-001
Alarms: Fetal Heart Rate Alarms
Fetal Heart Rate Alarms
FHR Patient Alarms
A fetal heart rate
patient
alarm occurs when any fetal heart rate falls outside of the
pre-defined alarm limits—greater than the high limit setting or less than the low
limit setting.
The FHR alarm function can be completely disabled from the password-protected
Install Options
service screen. For this change to take effect, you must cycle power.
Refer to the “250/250cx Series Monitor Service Manual” for more information.
NOTE:
The re-alarm time does not apply to FHR alarms—only MECG and
MSpO
2
alarms. FHR values are not configurable.
Active Patient Alarm
A patient alarm is indicated both visually and audibly. The visual indication is
provided by flashing the affected FHR numeric. The audio alarm consists of
alternating high/low tones.
Resolved Patient Alarm
Resolved FHR alarms function differently than other alarms with a 250cx Series
Monitor:
Resolved, Unsilenced FHR Patient Alarm:
You must acknowledge an FHR
patient alarm—even if the condition has already been resolved. The visual and
audible indications remain present until you press the
Alarm Silence
button.
This ensures that a clinician is aware that an alarm occurred. You may hear this
type of alarm described as
latching.
Resolved, Silenced FHR Patient Alarm:
If you have already silenced an FHR
patient alarm, the visual indications disappear automatically.
By comparison, the visual and audible indications for a
maternal
patient alarm
automatically disappear as soon as the condition is resolved—whether or not you
have acknowledged the alarm.
FHR Signal Quality Alarms
A fetal heart rate signal quality alarm occurs if the monitor is unable to detect an
acceptable FHR signal.
Active Signal Quality Alarm
A signal quality alarm is indicated both visually and audibly. The visual indication is
provided by flashing the FHR numeric (if available) or flashing dashes “– – –” in
place of the FHR numeric. The audio alarm consists of alternating high/low tones.
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......