Revision C
250cx Series Maternal/Fetal Monitor
11-5
2036946-001
Recorder Modes: Functionality with a QS System
CAUTION
DATA STORAGE—Stored maternal vital signs history data is
erased when you turn the
monitor off
. Therefore, for intermittent
monitoring, it is recommended that you leave the
monitor on
, but
turn the
recorder off.
Refer to Chapter 12, “Maternal Vital Signs
History” for more information.
Functionality with a QS System
Users of a Quantitative Sentinel (QS) System (Software Version 4.0.3.0 or earlier)
should be aware of the following items when using the monitor’s maternal-only
mode:
Paper Versus Electronic Strip Charts
As described earlier, the monitor’s maternal-only mode acts as a “paper saver”
turning the strip chart recorder on and off as needed. However, the QS System
overrides the maternal-only mode by storing the entire patient record. In other
words, the
electronic
strip chart is retained as if the monitor’s recorder were left on
Changing Recorder Modes
From
To
Button
Off
On
Briefly press once; or press and hold 3 seconds.
Off
Maternal-Only
Briefly press twice.
On
Maternal-Only
Briefly press once.
On
Off
Press and hold 3 seconds.
1
Maternal-Only
On
Briefly press once.
Maternal-Only
Off
Press and hold 3 seconds.
1
1
A confirmation tone sounds to indicate the recorder has been turned off.
Recorder Mode Audiovisual Indicator Status
Recorder Mode
Record Indicator
Audio Indicator
On
continuously illuminated
Off
Maternal-Only
lights three short flashes every
5 seconds
Off
Off
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......