Revision C
250cx Series Maternal/Fetal Monitor
D-7
2036946-001
Frequently Asked Questions:
How do I change the Maternal ECG lead
waveform being displayed?
Option 1
1. Rotate the Trim Knob to highlight the legend for MHR/P. (This legend is in the
approximate center of the display and reads MECG, Pulse or INOP, according to the
settings that are currently enabled.)
2. Once the MECG legend is highlighted, press the Trim Knob. The display changes to
show the MHR/P Setup screen.
3. Now rotate the Trim Knob to highlight the Source: setting. (It will read Auto, MSpO
2
,
MECG or NIBP.)
4. If the Source: setting is not MECG, press the Trim Knob again. The current setting
displays in blinking inverse video.
5. Now rotate the Trim Knob to change the current setting to MECG.
6. Once you set the desired source parameter, press the Trim Knob again to save the
value. The current value setting stops blinking.
7. Now rotate the Trim Knob to select (highlight) the MECG Lead: setting. This setting is
slightly above vertical center towards the right side of the screen and indicates Lead
I, II or III.
8. Now press the Trim Knob again. The current setting displays in blinking inverse
video.
9. Now rotate the Trim Knob to change the current setting to the desired lead selection.
10. Once you set the desired source parameter, press the Trim Knob to confirm your
selection. The current value setting stops blinking.
11. Now rotate the Trim Knob to select (highlight) the Exit item on the bottom menu.
12. Press the Trim Knob again to return to the main monitoring display screen.
Option 2
1. If the MHR/P Source setting is set to Auto, and MECG is currently enabled (plugged
in) OR MHR/P Source is set to MECG, then, from the main monitor screen, rotate the
Trim Knob to highlight the selected Lead setting for MECG. (This setting is slightly
below vertical center, on the right hand side of the display. It will indicate I, II or III.)
2. Once the MECG Lead setting legend is highlighted, press the Trim Knob. The
current setting displays in blinking inverse video.
3. Now rotate the Trim Knob to change the current setting to MECG Lead setting I, II or
III.
4. Once you set the desired source parameter, press the Trim Knob to confirm your
selection. The current value setting stops blinking and the waveform should reflect
the selected lead.
How do I change the waveform vertical
scale on the display?
1. To change the waveform scale, the current selected waveform must be MECG or
FECG. The waveform scale cannot be altered when you view MSpO
2
plethsysmograph waveforms which auto-scale.
2. Rotate the Trim Knob to highlight the scale factor setting for the waveform display.
(This setting is slightly below vertical center, on the right hand side of the display,
below the horizontal speed indication of 25 mm/s. The vertical scale indicates one of
the following: Auto, 0.25x, 0.5x, 1x, 2x, or 4x.
3. Once the MECG scale setting is highlighted, press the Trim Knob. The current
setting displays in blinking inverse video.
4. Now rotate the Trim Knob to change the current setting to Auto, 0.25x, 0.5x, 1x, 2x,
or 4x.
5. Once you set the desired scale, press the Trim Knob to confirm your selection. The
current value setting stops blinking, and the waveform should reflect the selected
lead.
Question
Answer
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......