8-8
250cx Series Maternal/Fetal Monitor
Revision C
2036946-001
Maternal Non-Invasive Blood Pressure Monitoring: NIBP Monitoring
NIBP Done Volume
This field sets the volume of the sound emitted at the completion of each blood
pressure determination. As you adjust the volume, a sample tone sounds.
Alarms
These fields adjust the high and low alarm limits for maternal systolic, diastolic, and
mean arterial pressures, as well as for MHR/P—in increments of 5 mmHg (0.7 kPa)
or 5 bpm. The selectable values are listed in “Maternal
page 8-7. The factory default settings are listed in Appendix A, “Factory Defaults”.
Alarm Volume
This field controls alarm volume for all maternal alarms.
NIBP Monitoring
Checklist
1.
The NIBP hose is securely inserted into the
NIBP
connector on the monitor.
2.
A cuff appropriate for the limb size has been selected.
3.
Cuff is properly placed on patient and connected to the NIBP hose.
4.
Tubes between the cuff and the monitor are not kinked or blocked.
Patient Preparation
Cuff selection and application are important. Inappropriate selection or improper
application of the cuff will result in erroneous measurements.
WARNING
The system is designed for use only with dual-hose cuffs and
tubing.
Do not place the cuff on a limb being used for A-V fistulas,
intravenous infusion or on any area where circulation is
compromised or has the potential to be compromised.
1.
Connect the air hose to the
NIBP
port on the front of the monitor. Make sure that
the hose is not kinked or compressed.
2.
Choose the appropriate blood pressure measurement site. Because normative
values are generally based on this site and as a matter of convenience, the upper
arm is preferred. When upper arm size or shape or the patient’s clinical
condition or other factors prohibit use of the upper arm, the clinician must plan
patient care accordingly, taking into account the patient’s cardiovascular status
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......