Service and Maintenance
5-13
December 2004
The Sensor Enclosure
Those components associated with the oxygen measurement process
are located in the sensor enclosure. This section describes the proper
procedures for replacing these parts.
!WARNING!
There are extremely hot surfaces in the sensor enclosure.
If these surfaces are touched, serious burns could result.
Always wear heat-resistant gloves while performing
these procedures.
IMPORTANT:
These symbols indicate Caution - dangerously hot
surfaces and risk of electric shock, respectively:
Caution!
Rapid cooling of the oxygen sensor will result in cracks in
the ceramic that will render the sensor unusable. Before
opening the door to the sensor enclosure, allow about two
hours after the system is powered down for the sensor to
cool slowly.
Proceed to the appropriate section to replace the desired component.