Setup and Calibration
Setting the Zero Trim
Using the procedure under “Testing the Analog Output” on page
4-21, set the
Test Percent
to 0.
At the analog output setup menu, proceed as follows:
Note:
The
[Aout Type], [Zero Setpoint]
and
[Span Trim]
options are also
available in the above list box.
At the next prompt, enter the desired value for the zero trim.
The meter now automatically returns to the “Analog Output” prompt.
Using the procedure under “Testing the Analog Output” on page
4-21, reset the
Test Percent
to its normal value.
Either press
[ESC]
until you exit the
User Program
or proceed to the
appropriate section to continue setting up the analog output.
User Aout
Use the [
] and [
] keys to
select
[Zero Trim]
, and press
[ENTER]
.
Span Setpoint
Test
Zero Trim
POWER
ESC
ENTER
User Aout Zero-Trim
Use [
] and [
] to increment
the value. Press
[ENTER]
when
done.
Zero Trim
s
0
t
POWER
ESC
ENTER
www.DataSheet4U.com
www.DataSheet4U.com
Summary of Contents for CGA 351
Page 7: ...Chapter 1 www DataSheet4U com www DataSheet4U com ...
Page 15: ...Chapter 2 www DataSheet4U com www DataSheet4U com ...
Page 27: ...Chapter 3 www DataSheet4U com www DataSheet4U com ...
Page 38: ...Chapter 4 www DataSheet4U com www DataSheet4U com ...
Page 77: ...Chapter 5 www DataSheet4U com www DataSheet4U com ...
Page 104: ...Chapter 6 www DataSheet4U com www DataSheet4U com ...
Page 109: ...Appendix A www DataSheet4U com www DataSheet4U com ...
Page 115: ...Appendix B www DataSheet4U com www DataSheet4U com ...
Page 118: ...Appendix C www DataSheet4U com www DataSheet4U com ...