Setup and Calibration
4-7
December 2004
Selecting the Operating
Mode
Enter the alarm setup menu, as described on page 4-2, and proceed as
follows:
Note:
The
[Trip Type]
and
[Trip Value]
options are also available in
the above list box.
At the next prompt, enter the desired operating mode.
Note:
In fail-safe mode, the alarm relay is constantly energized
during normal operation and is de-energized when the trip
point is crossed. In normal (non fail-safe) operation, the
alarm relay is energized only when the trip point is crossed.
The meter now automatically returns to the “Alarm-Features”
prompt. Press
[ESC]
until you exit the
User Program
.
Alarms..Alarm-Feat
Use the [
] and [
] keys to
select
[Fail-Safe]
, and press
[ENTER]
.
Deadband Value
Test
Fail-Safe
POWER
ESC
ENTER
Alarms..Fail-Safe
Use the [
] and [
] keys to
select the desired operating
mode and press
[ENTER]
.
Non Fail-Safe
Fail-Safe
POWER
ESC
ENTER
www.DataSheet4U.com
www.DataSheet4U.com
Summary of Contents for CGA 351
Page 7: ...Chapter 1 www DataSheet4U com www DataSheet4U com ...
Page 15: ...Chapter 2 www DataSheet4U com www DataSheet4U com ...
Page 27: ...Chapter 3 www DataSheet4U com www DataSheet4U com ...
Page 38: ...Chapter 4 www DataSheet4U com www DataSheet4U com ...
Page 77: ...Chapter 5 www DataSheet4U com www DataSheet4U com ...
Page 104: ...Chapter 6 www DataSheet4U com www DataSheet4U com ...
Page 109: ...Appendix A www DataSheet4U com www DataSheet4U com ...
Page 115: ...Appendix B www DataSheet4U com www DataSheet4U com ...
Page 118: ...Appendix C www DataSheet4U com www DataSheet4U com ...