General Information
The Sensor Enclosure
The sensor enclosure, which is shown in Figure 1-2 below, houses the
sample measurement components. The functions of these components
are as follows:
•
a zirconium oxide (zirconia)
oxygen sensor
converts the oxygen
concentration of the sample gas into a mV output signal.
•
a
sensor furnace
heats the oxygen sensor and the sample gas to the
optimum temperature of 700°C.
•
a
thermocouple
precisely regulates the temperature of the oxygen
sensor and the sample gas.
•
an inlet
needle valve
and an outlet
flowmeter
adjust the sample gas
flow to 1,000 ± 150 cc/min (2.1 ± 0.3 SCFH).
Figure 1-2: Sensor Enclosure
Furnace Bracket
Flowmeter
Thermocouple
Needle Valve
Sensor Furnace
Oxygen Sensor
Manifold
Electrical Conduit
Flowmeter
Furnace Bracket
Flowmeter
Thermocouple
Needle Valve
Sensor Furnace
Oxygen Sensor
Manifold
Electrical Conduit
www.DataSheet4U.com
www.DataSheet4U.com
Summary of Contents for CGA 351
Page 7: ...Chapter 1 www DataSheet4U com www DataSheet4U com ...
Page 15: ...Chapter 2 www DataSheet4U com www DataSheet4U com ...
Page 27: ...Chapter 3 www DataSheet4U com www DataSheet4U com ...
Page 38: ...Chapter 4 www DataSheet4U com www DataSheet4U com ...
Page 77: ...Chapter 5 www DataSheet4U com www DataSheet4U com ...
Page 104: ...Chapter 6 www DataSheet4U com www DataSheet4U com ...
Page 109: ...Appendix A www DataSheet4U com www DataSheet4U com ...
Page 115: ...Appendix B www DataSheet4U com www DataSheet4U com ...
Page 118: ...Appendix C www DataSheet4U com www DataSheet4U com ...