Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 5-23
Country/
Region
MODE
(Dose measured at 30 cm above the Image Intensifier)
Fluoro
kVp: 40 to 110
mA: 0.1 to 4
Pulsed Fluoro
kVp: 40 to 110
mA: 0.1 to 4
High Level
Fluoro
kVp: 40 to 110
mA: 0.2 to 12
Pulsed High
Level Fluoro
kVP:40 to 110
mA: 0.2 to 12
Digital Spot
kVp: 40 to 110
mA: 0.2 to 16
Film
kVp:40 to 110
mA: 20
mAs:1 to 80
Taiwan
(Republic of
China)
<87 mGy/min <87 mGy/min <100 mGy/min <100 mGy/min <2mGy
<20mGy
Thailand
<88 mGy/min <88 mGy/min <100 mGy/min <100 mGy/min <2mGy
<20mGy
Trinidad &
Tobago
<88 mGy/min <88 mGy/min <100 mGy/min <100 mGy/min <2mGy
<20mGy
Tunisia
<88 mGy/min <88 mGy/min <100 mGy/min <100 mGy/min <2mGy
<20mGy
Turkey
<88 mGy/min <88 mGy/min <100 mGy/min <100 mGy/min <2mGy
<20mGy
United Arab
Emirates
<88 mGy/min <88 mGy/min <100 mGy/min <100 mGy/min <2mGy
<20mGy
USA
<88 mGy/min <88 mGy/min <100 mGy/min <100 mGy/min <2mGy
<20mGy
United
Kingdom
<88 mGy/min <88 mGy/min <100 mGy/min <100 mGy/min <2mGy
<20mGy
Uruguay
<88 mGy/min <88 mGy/min <100 mGy/min <100 mGy/min <2mGy
<20mGy
Venezuela
<88 mGy/min <88 mGy/min <100 mGy/min <100 mGy/min <2mGy
<20mGy
Vietnam
<88 mGy/min <88 mGy/min <100 mGy/min <100 mGy/min <2mGy
<20mGy
The highest value for the Low Dose imaging
doesn’t exceed 50% of the
highest value for the Normal Fluoro.
The highest
value for the Low Level Pulsed Fluoro doesn’t exceed 50% of the
highest value for the Pulsed
Normal Fluoro.
Summary of Contents for Brivo OEC 715
Page 9: ...Chapter1 Introduction and Safety...
Page 34: ...Chapter2 System Overview...
Page 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Page 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Page 40: ...Chapter3 Mechanical Positioning...
Page 52: ...Chapter4 Start Up and Operating Control...
Page 77: ...Chapter5 Usual Operations...
Page 108: ...Chapter6 SpecialApplications...
Page 139: ...Chapter7 Radiographic Film...
Page 142: ...Chapter8 System Setup...
Page 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Page 150: ...Chapter9 Maintenance...
Page 160: ...Chapter10 Display Messages...
Page 162: ...Chapter11 Labels and Symbols...
Page 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Page 179: ...Chapter12 Technical Reference...
Page 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Page 206: ...Technical Reference Page 12 28 12 11 System Block...
Page 209: ......