![GE Brivo OEC 715 Operator'S Manual Download Page 193](http://html1.mh-extra.com/html/ge/brivo-oec-715/brivo-oec-715_operators-manual_4193902193.webp)
Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 12-15
12.5.
Scatter Radiation
Isokerma Map Test
This test method is according to clause 203.13.4 in EN 60601-2-43:2010.
The measurement is performed at 2 planes at typical height: 1m & 1.5m from the ground.
The scatter radiation is measured in 2 layouts as vertical and lateral configuration.
For the vertical configuration, the table is at height 1m. I.I. is above the tube.
For the lateral configuration, the table is at height 1m. The center of I.I. is at same altitude as the center of
phantom. The distance from the surface of I.I. to center of phantom is 40cm.
The layout is as below:
Summary of Contents for Brivo OEC 715
Page 9: ...Chapter1 Introduction and Safety...
Page 34: ...Chapter2 System Overview...
Page 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Page 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Page 40: ...Chapter3 Mechanical Positioning...
Page 52: ...Chapter4 Start Up and Operating Control...
Page 77: ...Chapter5 Usual Operations...
Page 108: ...Chapter6 SpecialApplications...
Page 139: ...Chapter7 Radiographic Film...
Page 142: ...Chapter8 System Setup...
Page 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Page 150: ...Chapter9 Maintenance...
Page 160: ...Chapter10 Display Messages...
Page 162: ...Chapter11 Labels and Symbols...
Page 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Page 179: ...Chapter12 Technical Reference...
Page 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Page 206: ...Technical Reference Page 12 28 12 11 System Block...
Page 209: ......