Technical Reference
Page 12-20
Vertical Plane
The following illustration shows measurement points in a plane through the beam axis at a height
indicated by reference point C1, which represents the 30-inch SSD. The measurement points are 25 cm
apart. The table following the illustration shows expected scatter radiation at these points.
Measurement
Point
Dose Rate
LOHMANN X-ray tube
T1
1.56 mGy/hr
mGy/hr
T2
0.60
mGy/hr
T3
0.28 mGy/hr
C1
0.01 Gy/min
B1
Too close to phantom for measurement
B2
2.03 mGy/hr
mGy/hr
B3
0.76 mGy/hr
mGy/hr
Summary of Contents for Brivo OEC 715
Page 9: ...Chapter1 Introduction and Safety...
Page 34: ...Chapter2 System Overview...
Page 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Page 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Page 40: ...Chapter3 Mechanical Positioning...
Page 52: ...Chapter4 Start Up and Operating Control...
Page 77: ...Chapter5 Usual Operations...
Page 108: ...Chapter6 SpecialApplications...
Page 139: ...Chapter7 Radiographic Film...
Page 142: ...Chapter8 System Setup...
Page 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Page 150: ...Chapter9 Maintenance...
Page 160: ...Chapter10 Display Messages...
Page 162: ...Chapter11 Labels and Symbols...
Page 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Page 179: ...Chapter12 Technical Reference...
Page 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Page 206: ...Technical Reference Page 12 28 12 11 System Block...
Page 209: ......