Introduction and Safety
Page 1-20
Guidance and manufacturer’s declaration
- Electromagnetic Immunity
The
Brivo OEC 715/785/865
Mobile C-Arm X-Ray Product is intended to use in the specified
electromagnetic environment. The purchaser or user of the
Brivo OEC 715/785/865
Mobile C-Arm X-Ray
Product should assure that it is used in an electromagnetic environment as described below:
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms , 6Vrms
at ISM band
150GHz to
80MHz
3V/m 80MHz to
2.5GHz (Ed3.0)
3V/m 80MHz to
2.7GHz, 28V/m
(Ed4.0)
3Vrms , 6Vrms at
ISM band 150GHz
to 80MHz
3V/m 80MHz to
2.5GHz (Ed3.0)
3V/m 80MHz to
2.7GHz, 28V/m
(Ed4.0)
Portable and mobile RF communications
equipment are used no closer to any part of
the
Brivo OEC 715/785/865
Mobile C-Arm
X-Ray Product, including cables, than the
recommended separation distance calculated
from the equation appropriate for the
frequency of the transmitter.
Recommended separation distance
P
=
d
]
3
5
,
3
[
150 kHz to 80 MHz
P
=
d
]
3
5
,
3
[
80 MHz to 800 MHz
P
=
d
]
3
7
[
800 MHz to 2.5 GHz
The formulas are applicable for Ed3.0 only.
Note: P is the power rating of the transmitter
in watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
, are less than the compliance level in each
frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following symbol:
a: Field strengths from fixed transmitters, such as base stations for cellular telephones and land mobile radios, amateur
radio, AM and FM radio broadcast, and TV broadcast cannot be estimated accurately. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be performed. If the measured field
strength exceeds the RF compliance level above, observe the Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product to
verify normal operation in each use location. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the [EQUIPMENT and/or SYSTEM].
b: Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Note3: These are guidelines. Actual conditions may vary.
Summary of Contents for Brivo OEC 715
Page 9: ...Chapter1 Introduction and Safety...
Page 34: ...Chapter2 System Overview...
Page 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Page 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Page 40: ...Chapter3 Mechanical Positioning...
Page 52: ...Chapter4 Start Up and Operating Control...
Page 77: ...Chapter5 Usual Operations...
Page 108: ...Chapter6 SpecialApplications...
Page 139: ...Chapter7 Radiographic Film...
Page 142: ...Chapter8 System Setup...
Page 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Page 150: ...Chapter9 Maintenance...
Page 160: ...Chapter10 Display Messages...
Page 162: ...Chapter11 Labels and Symbols...
Page 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Page 179: ...Chapter12 Technical Reference...
Page 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Page 206: ...Technical Reference Page 12 28 12 11 System Block...
Page 209: ......