Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 12-3
12.2.3.
Power Requirements
Voltage: AC 100V/110V/120V|200V|220V/230V/240V
Frequency: 50Hz/60Hz
Rated Momentary Line Current (shorter than 5s):
•
30A (100V/110V/120V)
•
28A (200V)
•
25A (220V/230V/240V)
Rated Continuous Line Current:
•
20A (100V/110V/120V)
•
12A (200V)
•
10A (220V/230V/240V)
Maximum line impedance: 0.3
|0.6
|0.6
Max Line Regulation (Percent): 15% at maximum radiographic exposure.
Line regulation is based on measurements made at the input (primary winding) of the isolation transformer
under standby conditions and at maximum radiographic exposure by the following equation:
Line regulation = 100 (Vn - V1)/V1
Where Vn = no load voltage (standby), and V1 = loaded voltage (maximum radiographic exposure).
Technique factors that constitute the maximum line current condition is 110kV, 20mA.
16A power supply is required in China. And 20 A is required in Japan.
NOTE
Summary of Contents for Brivo OEC 715
Page 9: ...Chapter1 Introduction and Safety...
Page 34: ...Chapter2 System Overview...
Page 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Page 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Page 40: ...Chapter3 Mechanical Positioning...
Page 52: ...Chapter4 Start Up and Operating Control...
Page 77: ...Chapter5 Usual Operations...
Page 108: ...Chapter6 SpecialApplications...
Page 139: ...Chapter7 Radiographic Film...
Page 142: ...Chapter8 System Setup...
Page 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Page 150: ...Chapter9 Maintenance...
Page 160: ...Chapter10 Display Messages...
Page 162: ...Chapter11 Labels and Symbols...
Page 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Page 179: ...Chapter12 Technical Reference...
Page 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Page 206: ...Technical Reference Page 12 28 12 11 System Block...
Page 209: ......