Technical Reference
Page 12-6
12.3.
Specification of Key Components
12.3.1.
X-Ray Generator
•
Nominal electric power:
2.0kW (100kV, 20mA, 0.1s)
•
Maximum electric power:
Normal Fluoro
440W (110kV, 4mA)
High Level Fluoro
660W (110kV, 6mA; 55kV, 12mA)
Low Dose
220W (110kV, 2mA)
Digital Spot
1.76 kW (110kV, 16mA) @ 200V, 220V, 230V, 240V;
1.1 kW (110kV, 10mA) @ 100V, 110V, 120V
Film
2.2 kW (110kV, 20mA) @ 200V, 220V, 230V, 240V;
1.1 kW (110kV, 10mA) @ 100V, 110V, 120V.
•
High frequency generator:
40kHz
•
Voltage:
40kV~110kV
Normal Fluoro
When the X-Ray tube current reaches its highest value of 4mA, the
obtainable highest X-Ray tube voltage is 110kV.
High Level Fluoro
When the X-Ray tube current reaches its highest value of 12mA, the
obtainable highest X-Ray tube voltage is 55kV.
Low Dose
When the X-Ray tube current reaches its highest value of 2mA, the
obtainable highest X-Ray tube voltage is 110kV.
Digital Spot
When the X-Ray tube current reaches its highest value of 16mA, the
obtainable highest X-Ray tube voltage is 110kV @ 200V, 220V, 230V,
240V.
When the X-Ray tube current reaches its highest value of 10mA, the
obtainable highest X-Ray tube voltage is 110kV @ 100V, 110V, 120V.
Film
20mA, stationary @ 200V, 220V, 230V, 240V, the obtainable highest X-Ray
tube voltage is 110kV;
10mA, stationary @ 100V, 110V, 120V, the obtainable highest X-Ray tube
voltage is 110kV.
•
Current:
Normal Fluoro
0.1 mA~4 mA; when the X-Ray tube voltage reaches its highest value of
110kV, the obtainable highest X-Ray tube current is 4mA.
High Level Fluoro
0.2 mA~12 mA; when the X-Ray tube voltage reaches its highest value of
110kV, the obtainable highest X-Ray tube current is 6mA.
Low Dose
0.1 mA~2 mA; when the X-Ray tube voltage reaches its highest value of
110kV, the obtainable highest X-Ray tube current is 2mA.
Digital Spot
0.2 mA~16 mA; when the X-Ray tube voltage reaches its highest value of
110kV, the obtainable highest X-Ray tube current is 16mA @ 200V, 220V,
230V, 240V.
0.2 mA~10 mA; when the X-Ray tube voltage reaches its highest value of
Summary of Contents for Brivo OEC 715
Page 9: ...Chapter1 Introduction and Safety...
Page 34: ...Chapter2 System Overview...
Page 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Page 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Page 40: ...Chapter3 Mechanical Positioning...
Page 52: ...Chapter4 Start Up and Operating Control...
Page 77: ...Chapter5 Usual Operations...
Page 108: ...Chapter6 SpecialApplications...
Page 139: ...Chapter7 Radiographic Film...
Page 142: ...Chapter8 System Setup...
Page 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Page 150: ...Chapter9 Maintenance...
Page 160: ...Chapter10 Display Messages...
Page 162: ...Chapter11 Labels and Symbols...
Page 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Page 179: ...Chapter12 Technical Reference...
Page 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Page 206: ...Technical Reference Page 12 28 12 11 System Block...
Page 209: ......