Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
Page 12-19
Horizontal & Vertical Plane Test
This test is performed at fluoroscopic mode in 70kV, 3mA, without shielding.
Horizontal Plane
The following illustration shows measurement points in a plane perpendicular to the beam axis at a
height indicated by reference point C1. C1 is the 30-inch SSD, and is illustrated in the Vertical Plane
illustration, below. The measurement points are 25 cm apart. The table following the illustration shows
expected scatter radiation at each point.
Measurement
Point
Dose Rate
LOHMANN X-ray tube
H1
3.32 mGy/hr
H2
0.67 mGy/hr
H3
0.28 mGy/hr
H4
0.19 mGy/hr
S1
3.60 mGy/hr
S2
0.81 mGy/hr
S3
0.36 mGy/hr
S4
0.19 mGy/hr
Summary of Contents for Brivo OEC 715
Page 9: ...Chapter1 Introduction and Safety...
Page 34: ...Chapter2 System Overview...
Page 36: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 3 2 2 C Arm Components...
Page 38: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 2 5 2 3 Workstation Components...
Page 40: ...Chapter3 Mechanical Positioning...
Page 52: ...Chapter4 Start Up and Operating Control...
Page 77: ...Chapter5 Usual Operations...
Page 108: ...Chapter6 SpecialApplications...
Page 139: ...Chapter7 Radiographic Film...
Page 142: ...Chapter8 System Setup...
Page 149: ...System Setup Page 8 8 Click any mouse button to return to normal screen...
Page 150: ...Chapter9 Maintenance...
Page 160: ...Chapter10 Display Messages...
Page 162: ...Chapter11 Labels and Symbols...
Page 177: ...Labels and Symbols Page 11 16 42 12 15 29 40 15 13 10 18 39 31 27 24 23 37 38...
Page 179: ...Chapter12 Technical Reference...
Page 183: ...Brivo OEC 715 785 865 Mobile C Arm X Ray Product Page 12 5 Workstation Weight 170 10kg...
Page 206: ...Technical Reference Page 12 28 12 11 System Block...
Page 209: ......