Action
Step
Tap Diagnostics.
3
Result: The Diagnostics screen opens.
Tap Set baseline to capture a new reference value for the test.
4
Tap Dark current test.
5
The result is displayed as Pass or Fail:
•
Pass: the UV Monitor is working as it should.
•
Fail: replace the UV Monitor.
Check for presence of stray light
Follow the instructions below to perform a Stray light test.
Action
Step
Flush and fill the UV flow cell with 15% acetone using a syringe. Make sure
that there are no air bubbles in the syringe.
1
ÄKTA start Maintenance Manual 29-0603-08 AC
67
5 Service and settings
5.4 UV Monitor
5.4.2 Diagnostics