5.14.4
Switch valve timing
Instructions
Follow the instructions below to set the Switch valve timing. For more details, refer to
ÄKTA start Operating Instructions section Switch valve timing.
Action
Step
In the Settings and service screen, tap System.
1
Result: The following screen opens.
In the System screen, tap Switch valve timing.
2
Result: The following screen opens.
Tap the radio button to select Switch valve timing B (switch time 5 sec.).
3
Tap Save to save the timing.
Perform Gradient run, either by performing System performance method
or manually set the B concentration (Buffer valve) to 50%. Observe for the
fluctuations.
4
ÄKTA start Maintenance Manual 29-0603-08 AC
125
5 Service and settings
5.14 System
5.14.4 Switch valve timing