Action
Step
In the Settings and service screen, tap Pressure sensor.
2
Result: The Pressure sensor screen opens.
In the Pressure sensor screen, tap Zero offset.
3
Result: A Message screen opens.
Make sure that there is no back pressure in the system, and then tap OK.
102
ÄKTA start Maintenance Manual 29-0603-08 AC
5 Service and settings
5.11 Pressure sensor