Action
Step
If wavy gradients still are obtained, or if fluctuations on step gradient levels
are large, then select Advanced timing.
5
Result: The following screen opens.
Set switch valve time in the range of 3.0 to 5.0 sec (0.1 sec increments) by
pressing the up/down arrows.
Tap Save to save the optimized timing.
6
126
ÄKTA start Maintenance Manual 29-0603-08 AC
5 Service and settings
5.14 System
5.14.4 Switch valve timing