7.10
Conductivity Monitor
Required tools
Dimension
Tool
T20
Torx screwdriver
Instructions
Follow the instructions below to remove and replace the Conductivity Monitor.
WARNING
Disconnect power. Always switch off power to the instrument be-
fore replacing any component on the instrument or cleaning the
instrument, unless stated otherwise in the user documentation.
If required, twist the connector to bring the connector out of the instrument
hole, before disassembly the Conductivity Monitor.
Note:
Action
Step
Switch off the instrument by pressing the Power Switch to the O position.
1
Remove the inlet and outlet tubing from the ports.
2
182
ÄKTA start Maintenance Manual 29-0603-08 AC
7 Disassembly
7.10 Conductivity Monitor