Action
Step
Remove the Pump.
6
Note:
Make sure that the pump cable is placed away from the fan, on the left side
of the instrument after removing the Pump. This safety precautions is to ensure
that the cable does not get stuck in the ventilation fan.
Replace with a new Pump, install the new Pump in reverse order.
7
Place the Pump tubing between the rollers and the track of the new Pump.
8
Switch on the instrument by pressing the Power Switch to the l position.
9
In the Pump screen, tap Diagnostics and then reset the number of hours
of Pump run to 0.
10
ÄKTA start Maintenance Manual 29-0603-08 AC
171
7 Disassembly
7.6 Pump