7.6
Pump
Required tools
Dimension
Tool
T20
Torx driver
Instructions
Follow the instructions below to remove and replace the Pump.
WARNING
Disconnect power. Always switch off power to the instrument be-
fore replacing any component on the instrument or cleaning the
instrument, unless stated otherwise in the user documentation.
NOTICE
Keep the pump cover open when not using the system. Open the
peristaltic pump cover after you switch off the equipment. This re-
duces the risk of shortening the life time of the pump tubing.
Action
Step
Switch off the instrument by pressing the Power Switch to the O position.
1
ÄKTA start Maintenance Manual 29-0603-08 AC
167
7 Disassembly
7.6 Pump