5.14.3
Delay volume setting
Instructions
Follow the instructions below to set the delay volume.
Action
Step
In the System screen, tap Delay volume setting.
1
Result: The following screen opens.
Enter the internal diameter (ID) and length of the tubing from the Outlet
valve to Frac30 in the respective fields, and then tap Save.
2
Result: The total delay volume from the UV Monitor to Frac30 is displayed
in the Total Delay volume field.
Note:
The delay volume from the UV Monitor to the Outlet valve is constant (0.27
ml) in all ÄKTA start instruments.
Delay volume setting needs to be set when the tube between Outlet valve
and Frac30 has been replaced.
Note:
124
ÄKTA start Maintenance Manual 29-0603-08 AC
5 Service and settings
5.14 System
5.14.3 Delay volume setting