15
gbo Medizintechnik
AG 2009
Version 1.0
4 Troubleshouting
Each problem that occurs during operation of the unit will be shown in a message window on top of
the display and also signalised by an acoustic tone. Most of the problems can be solved by the
instructions displayed.
In general:
1. The malfunction is shown on the display.
2. An acoustic signal is heard.
3. Follow the instructions on the display.
Suggestions:
•
Turn off and turn on the unit.
•
A full selftest of all device functions is performed and the unit is reinitialized. If the unit
does not come up to normal state, please call your local service.
4.1
Further errors
symptom
cause and action
Device cannot be switched on,
no display contents shown.
Please check the mains plug and sockets. If
necessary; contact your service agents or the
manufacturer.
No acoustic signal is heard.
(End of therapy…)
Please check the settings of the acoustic tones in
the menu. The volume must be greater than 0.
Please contact your service agent or the manufacturer if the problems cannot be solved by the
measures mentioned above.
Please note that the unit must be placed on a plane horizontal surface. The device should neither be
placed in front of radiators nor should it be covered with pillows or blankets while in operation. Do
not cover the ventilation slots on the bottom of the unit either.
4.2 Error
codes
Error
code
Error message
Probable cause
Possible remedy
30033 Error during therapy: High temperature
at heat sink
Ultrasound probe defective
Power amplifier defective
Let the device cool down.
Check with another Ultrasound probe.
If the problem still exists change
ultrasound amplifier board.
30034 Error during therapy: High current
(Software)
Ultrasound probe defective
Check with another Ultrasound probe.
30035 Error during therapy: High current
(Hardware)
Ultrasound probe defective
Check with another Ultrasound probe.