37
gbo Medizintechnik
AG 2009
Version 1.0
7.1 Current
types
The
Stereodynator
offers a large variety of current types for therapy. For details see the direct help
menu on the display of the device. The following current types are available as therapeutic
programs with the
Stereodynator:
Current Graphics Pulse
duration
T in ms
Pause
duration
R in ms
D.C.
portio
n in
%
Frequency
in Hz
Max.
output
Voltage
V
peak
Max.
output
current
mA
peak
Diadynamic
currents
CP (Courtes
périodes)
47.25
50
/
100
200
80
CP-id (MF-part
lowered)
45
50
/
100
200
80
DF (Diphasé
fixe)
63
100
200
80
LP (Longues
périodes)
46
50/100
200
80
MF (Monophasé
fixe)
31.5
50
200
80
MM
(Monophasé
modulé)
20
100
200
80
RS (Rhythme
syncope)
15.25
50
200
80
Pulsed
currents
Galvanic current
100
0 70
70’
High voltage
0.02 – 0.06 5 ms – 1 s
0
1 – 200
200
80
Impulse
galvanization IG
30
30.0 50.0 18
12.5 200
80
Impulse
galvanization IG
50
50.0 70.0 0.3
8.3 200
80
Frequency
modulation
FM / 7 – 14 Hz
1
70.9 – 142.4 0.4
7 - 14
200
80
Micro current
biphasic
2,5 - 166
2,5 - 166
0
3 – 200
200
80