36
gbo Medizintechnik
AG 2009
Version 1.0
7 Technical
Data
Mains voltage and frequency: 100 – 240 V, 50-60 Hz
Power consumption*: max.
200VA
Mains fuse :
T 5A L 230V
Output current:
max. 125 mA
peak
Output voltage:
max. 200 V
MDD device class:
IIa
Protection against ingress of: I acc. to IEC 601 / VDE 0750
Protection class:
BF acc. to IEC 601
Protection degree:
IP X0
Dimensions*:
max. 35 cm x 29 cm x 35 cm (W x H x D)
Weight*:
max. 13 kg without accessories
Color:
Aluminium nature anodized and grey RAL 7016
Display:
15“ TFT LCD Touchscreen
Battery: CR
2032
environmental conditions:
operation of the
device:
temperature range +10 °C ... +40 °C
relative humidity of air 30 ... 75 %
transport
and
storage:
temperature range +5 °C ... +50 °C
relative humidity of air < 90 %, non condensing
Current types:
See chapter 7.1
Carrier
Frequency
2kHz – 10kHz Sinus wave
Low
frequency 0.1Hz
– 200Hz
Accuracy
±
10%
Therap.ultrasound module:
Continuous and pulse mode 1:1, 1:2, 1:5, 1:10, 1:20
Ultrasound frequency:
1 & 3 MHz
Ultrasound power:
max. 3 W per 1 cm
2
radiating surface at 1 MHz
max. 1.5 W per 1 cm
2
radiating surface at 3 MHz
Ultrasound probe: 2.5
cm
2
and 5 cm
2
radiating surface
Accuracy
±
10%
Suction application
Suction:
At least 180 mbar in maximum position
Suction massage frequency:
0 - 30 pulses per minute
*) dependent on the device type and/or whether the device is equipped with therapeutic ultrasound modules and suction application
modules.
By request of technical personnel gbo Medizintechnik can offer spare part lists and circuit diagrams.
gbo Medizintechnik AG reserves the right to modify design and specifications without prior notice
.